Milner, Thomas E.847530582008-08-282017-05-112008-08-282017-05-112006http://hdl.handle.net/2152/2595textDispersion caused by the refractive index variation over a spectral range is an important characteristic to identify the structure and composition of materials. This research reports on work to obtain dispersion information using both time and spectral domain optical coherence tomography. To process time-frequency data, a non-uniform Fourier transformation is applied to remove the resolved non-uniform frequency sampling. Analysis of the spectral phase function in the optical frequency domain is applied to measure the dispersion. First, this research experimented with water (H2O) to measure dispersion. The measured dispersion of water is compared with known data to confirm the methodology. Second, the concentration of a glucose solution was estimated by analyzing the spectral phase function. The result showed that this method can provide an ability to measure glucose concentration with high sensitivity 0.54 mM. In conclusion, this method can be implemented to monitor sample constituents and to compensate for material dispersion.electronicengCopyright is held by the author. Presentation of this material on the Libraries' web site by University Libraries, The University of Texas at Austin was made possible under a limited license grant from the author who has retained all copyrights in the works.Diagnostic imagingDispersionDispersion in biomedical optical imaging systemsThesis